Gasanalyse
Rapidox 7100 Biogas Analysator
The Rapidox 7100 Biogas Analyser is a high specification benchtop or rack mountable instrument designed for the on-line analysis, control, monitoring and calculation of calorific value for biogas produced from the anaerobic digestion of organic materials.
Please contact us for further versions and options
The Rapidox 7100 Biogas Analyser can be used in a laboratory or on-site at a fertilisation plant, landfill site, farm or sewage plant to measure both gas content (CH4 / H2S / O2 / CO2) and energy content (calorific value, CV) for biogas produced in a digester from biomass; up to six gases can be measured simultaneously meaning that other gases can be analysed as well, if required. The Rapidox 7100 Biogas Analyser is available as either a benchtop version for the laboratory or in a 19" rack mountable case which can be housed in a suitable enclosure for use in an external or hazardous area.
An optional pump enables two modes of operation; for samples that are taken from a gas source at atmospheric pressure or below, the pump is activated to draw a sample through the analyser and when sampling from a source at a greater atmospheric pressure, the pump can be deactivated, allowing the gas to flow directly through the analyser. Gas flow is regulated manually via a rotary knob on the fascia and displayed electronically on the screen.
CH4 (methane) |
0-100%, IR sensor (other ranges available) |
CO2 (carbon dioxide) |
0-100%, IR sensor (other ranges available) |
O2 (oxygen) |
0-30%, electrochemical sensor |
H2S (hydrogen sulphide) |
0-2000ppm or 0-10,000ppm / 0-1% (other ranges available) |
Other gases |
Contact us for further information |
Normal Operating Conditions |
Temperature 0°C to 40°C / Humidity 10 - 90% RH / Pressure 900 to 1100 mbar absolute. |
Sampling |
Fixed or continuous sampling modes |
Warm-up Time |
3-4 minutes at 20°C |
Voltage Outputs |
0-10V, user programmable |
Current Outputs |
4-20mA linear, user programmable |
Digital Outputs |
RS232 (RS485 option available) Data streamed on demand. Modbus RTU/Ethernet |
Data Output |
Excel compatible data via USB memory stick |
Alarms |
Relay circuits, user programmable |
Sample Connections |
6mm OD or 1/4" Swagelok fittings. Rear positioning |
Calibration |
Calibration by the user using zero and span gases An optional Auto-Cal system can be installed on certain models |
Display |
7" (180mm) full-colour LCD with touchscreen operation; resolution 0.01ppm or 0.01% |
Supply Voltage |
90-260 VAC, 50/60Hz |
Analyser Dimensions |